Velocity Edge DualX Servisní příručka Strana 129

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 400
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 128
Reference, Installation, and Operations Manual Section 5: Startup and meter configuration
3-9000-743 Rev U August 2014
Initial Ethernet communications connection 103
5.3.2 Initial Ethernet communications connection
For Ethernet communications, configure these data points (listed in the table below):
5.3.3 Ethernet communications connection
Table 5-3 Data Points for Ethernet Port Configuration
Daniel MeterLink
Display Name
Data Points, Options and Guidelines
IP Address Data points affected:
Eth1IPAddr
Options:
Enter a value (stored as a string)
Subnet mask Data points affected:
Eth1SubnetMask
Options:
Enter a value (stored as a string)
Default gateway Data points affected:
EthDfltGatewayAddr
Options:
Enter a value (stored as a string)
These instructions involve setting Mark III switches. Refer to Section 3.4 for further
information on switch settings.
Refer to Appendix D for communication troubleshooting information.
It is strongly recommended that the meter be configured using an independent (off-network)
single host computer. After configuration of the meter, the DHCP option should be disabled.
Zobrazit stránku 128
1 2 ... 124 125 126 127 128 129 130 131 132 133 134 ... 399 400

Komentáře k této Příručce

Žádné komentáře